pEPR Info file - v. 0.0.1 (2013-06-29) GENERAL Filename: Sample06-HYSCORE-20K-1sc-03 Date: 2013-08-30 Time start: 07:40:00 Time end: 15:17:00 Operator: Max Mustermann Experiment: HYSCORE Label: Sample 06 HYSCORE 03 Purpose: Sample 06 HYSCORE good resolution Spectrometer: Elexsys E680 Software: Xepr Runs: 1 SAMPLE Name: Bla 13CN immediately frozen Description: Bla 13CN immediately frozen (sample 06) Buffer: N/A Preparation: N/A Tube: 3,8 x 3,0 x 250 mm MAGNETIC FIELD Start: 348,30 mT Stop: 348,30 mT Step: 0 mT Sequence: none Controller: Bruker Power supply: Bruker ER 083 CS BRIDGE Model: Bruker TeraSpec Pulsed X-Band I.F. Bridge Attenuation: 05 dB Power: xx mW Detection: mixer Frequency counter: Bruker MW frequency: 9.69053 GHz MICROWAVE AMPLIFIER Type: TWT Model: Applied Systems Engineering 117X Power: 1 kW VIDEO AMPLIFIER Bandwidth: 25 MHz Amplification: 66 dB RECORDER Model: Bruker SpecJet Averages: 100 Time base: 4 ns Bandwidth: 25 MHz PROBEHEAD Type: dielectric ring Model: Bruker ER 4118X-MD5 Coupling: overcoupled TEMPERATURE Temperature: 20 K Controller: Oxford ITC 503 Cryostat: Oxford 935 Cryogen: LHe PULSE PROGRAM Pi-Pulse: 32 ns Sequence: pi/2-tau-pi/2-T1-pi-T2-pi/2-tau-acq SRT: 1000 us Scans: 4 AcquisitionMode: PulseSPEL PULSESPEL Program: HYSCORE Experiment: HYSCORE PhaseCycle: 4-step COMMENT tau = 200 ns T1 = T2 = 100:16:4196 ns pi/2 = 16 ns inversion pulse = 32 ns